Avoiding Exposure in Pregnancy

No law or professional standard requires that radiologists determine in advance whether a patient of childbearing-age is pregnant [1]. However, it is clearly good practice to implement the following guidelines:

Signs should be prominently displayed in all radiology departments asking each patient to notify a technologist or physician if she is, or thinks she could be, pregnant.
All technologists should ask women of childbearing-age if they might be pregnant prior to performing a radiologic procedure.
Radiology requisition forms filled out by referring physicians should include a section dealing with the possibility of pregnancy.
No radiological procedure involving exposure to the pelvis should be undertaken in a patient who declares she may be pregnant without consultation with a radiologist. The radiologist should discuss risks and benefits with the patient, and determine if it is appropriate to proceed, perform an alternative procedure, or delay the study to allow performance of a pregnancy test.

It should be noted that current recommendations do not recognize a safe period during the menstrual cycle, and so the concept of the "ten day rule" is obsolete. A patient who thinks she may be pregnant should be discussed with the referring physician, in order to determine the appropriate course of action (e.g., rescheduling after pregnancy testing, proceeding with the test after counseling, or changing to another modality).

Key point: It is the responsibility of the patient to disclose any possibility of pregnancy, although appropriate signage and questioning of all women of reproductive age is also critical. The supervising radiologist should discuss any cases of possible pregnancy with the referring physician.

Scanner for security an radiation


In our frightening new world, x-rays and CTs have become as important for airports as for hospitals.  Hand luggage is put through x-ray scanners; because one pass exposes your items to only a tiny fraction of the energy used in a chest x-ray, these scanners are safe for photographic film — and for the security personnel who use them all day long.  But checked baggage is passed through CT scanners that are far more powerful; they will damage your film but not your medications.

The new full-body scanners have raised concerns about privacy and safety.  The privacy issue is personal, but the safety issue is a scientific matter — and scientists agree that scanners are safe.  Two types are in use.  Millimeter wave scanners use radio waves to generate images, and they don’t expose travelers to any ionizing radiation.  Backscatter scanners do use low-intensity x-rays, but they bounce off the skin without penetrating the body.  They deliver only a tiny amount of radiation, about the same amount as you get in three to nine minutes of daily living, or about 1/1000 as much as an ordinary chest x-ray.  A person would need to have 2,500 to 5,000 backscatter scans a year to reach the established annual safety limit.

3D Mammograms May Actually TRIPLE Your Radiation Dose


Just how much more radiation are you getting with these new 3D mammograms? There are different estimates in circulation, partly because some folks are not including the radiation exposure from the standard mammogram into the total. When you add the 2D and 3D scans together, utilizing tomosynthesis at least doubles your radiation exposure, and some estimates have your exposure tripling, such as this 2011 study published in Radiology Today,2 which states that:

"Because the digital breast tomosynthesis (DBT) exam requires two additional exposures over a standard mammogram, the total radiation dose from the combined 2D and tomosynthesis examination is three times that of a standard mammogram."

The authors note that both scans (2D and 3D) are required because studies done prior to FDA approval failed to conclusively show better sensitivity of 3D alone, compared to 2D imaging.3 They also state that it’s too early to know how useful this new breast imaging technology will be. Some insurance companies—Aetna4 for one—have rejected it, citing “insufficient evidence of its effectiveness.”

The bottom line is, the industry reports tomosynthesis has 1.5 to 2 times the radiation dose. The FDA reports it has double the dose. And the radiologists, who are looking at total exposure, report triple the dose over conventional mammograms. You can see this comparison in a chart on page 8 of an FDA Executive Summary, “Average Glandular Dose per Breast.”5

Tomosynthesis is being touted by the industry as being particularly helpful for identifying cancer in women with dense breast tissue. However, these women already have a four to six times greater risk of developing breast cancer. Knowing that ionizing radiation is a direct cause of cancer, how can a test that triples your radiation exposure be of any benefit if you have potentially cancer-prone breasts to begin with?

Are there any possible dose reduction strategies for CT Colonography?

Yes. There are clear opportunities for dose reduction with almost any type of CT scan. Of importance are using the highest pitch and lowest tube current (mAs) consistent with acceptable images. Increasing the pitch from 1.0 to 2.0 usually reduces the dose by half. The patient must be carefully centred at the isocentre of the scanner. Tube current should be set at the lowest level that allows adequate visualization of the colonic wall , or tube current modulation (TCM), a technique that adapts tube current to the morphology of the patient, can be used to achieve dose reductions of about 35%.Most protocols use kVp between 110-120. A number of authors have developed innovative colonic phantoms to help optimize CTC protocols.

Ultra-low dose protocols result in an effective dose of 0.5 to 2 mSv using 10 mAs and 120 kVp These have been shown to be capable of a sensitivity of over 80% for polyps > 5 mm and specificity in excess of 95%. Imaging of polyps with doses as low as 0.05 mSv has been shown to be feasible. When compared with results from 8-12 mSv scans, the lesion detectability did not change significantly although image noise increased substantially. The latter problem can be ameliorated through use of noise reduction filters and image smoothing.This adds to the widely shared opinion that there is much scope for dose reduction, with some loss of image quality, but without significant reduction in lesion detectability.

Risks of a full-body CT scan?

The chance of a full-body CT scan causing a problem is small.

  • There is a chance of an allergic reaction to the dye (contrast material).
  • If you have diabetes or take metformin (Glucophage), the dye may cause problems. Your doctor will tell you when to stop taking metformin and when to start taking it again after the test so you will not have problems.
  • There is a small chance of developing cancer from having some types of CT scans. The chance is higher in children, young adults and people who have many radiation tests. If you are concerned about this risk, talk to your doctor about the amount of radiation this test may give you or your child and confirm that the test is needed.

The Risks of a Lumbar MRI Scan


Unlike X-rays and computed tomography (CT) scans, an MRI does not use radiation. It is considered a safer alternative for everyone, especially pregnant women. Although side effects do sometimes occur, they are extremely rare. There have been no documented side effects from the radio waves and magnets used in the scan to date.

Radiation Dosage and Decay


There are several methods to measure radiation; hence, there are several radiation units based on different radiation factors. Radiation units can measure radioactive decay, absorbed dosage, and human absorbed doses. Bq and Ci measure radioactive decay, while Gy and Rad measures absorbed doses. Sv and Rem measure absorbed doses in Gy and Rad equivalents. Rem takes into account different radiation types and the speed of particles. Below is a chart to help organize the different units:

Units for Radioactive Decay
Becquerel, BqMeasured in s-1, as disintegration per second
Curie, Ci
Measured as amount of decay at the same rate as 1 gram of radium
1 Ci = 3.70∙1010 Bq
Units for Absorbed Dose
Gray, Gy1 Gy deposits 1 Joule of energy per kilogram of matter
Rad1 rad = 0.01 Gy
Equivalent Doses
Sievert, Sv1Sv= 100 rem
Rem
1 rem = 1 rad∙Q
Q = 1 for X-rays, gamma rays, and beta particles
Q = 3 for slow neutrons
Q = 10 for protons and fast neutrons
Q = 20 for alpha particles
The most commonly used unit is the "rad," which stands for "radiation absorbed dose," and the "rem," which stands for "radiation equivalent for man." One rad corresponds to the absorption of 0.01 Joules of energy per kilogram of matter. Rem is the rad multiplied by the relative biological effectiveness, which is most often expressed as the variable "Q." The factor Q is used to take into account the different effects caused by different radiation.

Radiation and DNA


Radiation is simply a mechanism whereby energy passes through space. It takes the form of an electromagnetic wave, with the frequency of the electromagnetic wave determining its position in the electromagnetic spectrum. Low-frequency waves such as radio waves lie at one end of the spectrum and high-energy, high-frequency X-rays/Gamma rays at the other end. These high-frequency, high-energy waves are termed “ionizing” (as opposed to non-ionizing) radiation because they contain sufficient energy to displace an electron from its orbit around a nucleus. The most important consequence of this displaced electron on human tissue is the potential damage it can inflict on DNA, which may occur directly or indirectly. Direct damage occurs when the displaced electron hits and breaks a DNA strand. Indirect damage occurs when the electron reacts with a water molecule, creating a powerful hydroxyl radical which then damages the cell’s DNA.
Damage to a cell’s DNA in either of these ways can have several consequences. A single-strand DNA break is usually repaired appropriately by the cell with no subsequent deleterious sequelae. However, a break affecting both strands of DNA allows the potential for abnormal reconnection of the strands, which likely accounts for all the adverse biological effects ionizing radiation has on humans. First, DNA may rejoin itself incorrectly, rendering the cell nonviable with cell death taking place. Second, it may rejoin as a symmetrical translocation with the potential expression of an oncogene during division (and development of subsequent malignancy) or with abnormal division in gonads, giving rise to potential hereditary disorders.
Radiosensitivity is the probability of a cell, tissue, or organ suffering an effect per unit dose of radiation. Radiosensitivity is highest in cells which are highly mitotic or undifferentiated. For this reason the basal epidermis, bone marrow, thymus, gonads, and lens cells are highly radiosensitive. Muscle, bones, and nervous system tissues have a relative low radiosensitivity.

Aren't children more sensitive to radiation than adults?


Yes, because children are growing more rapidly, there are more cells dividing and a greater opportunity for radiation to disrupt the process. EPA's radiation protection standards take into account the differences in the sensitivity due to age and gender.

Fetuses are also highly sensitive to radiation. The resulting effects depend on which systems are developing at the time of exposure.

Prenatal Radiation Exposure



Rapidly proliferating and differentiating tissues are most sensitive to radiation damage. Consequently, radiation exposure can produce developmental problems, particularly in the developing brain, when an embryo/fetus is exposed prenatally.

The developmental conditions most commonly associated with prenatal radiation exposure include low birth weight, microcephaly, mental retardation, and other neurological problems. These effects are related to the developmental stage at which the exposure occurs. The threshold dose for developmental effects is approximately 10 rems.

The evidence that the developing embryo/fetus is more sensitive to radiation-induced cancer is inconclusive. But it is prudent to assume that there is some increased sensitivity.

What about the radiation dose in OPG?


Although the radiation dose associated with OPG and cephalometry is extremely low it is important that the procedure is performed using modern dose efficient equipment. In 2006 Barwon Health installed new digital OPG and Cephalometry systems that utilise direct radiography (DR) capture of the x-rays. This results in a potential dose reduction of 50% compared with conventional x-rays using either film or computed radiography (CR) phosphor plate technology. There are also specific paediatric dose reduction protocols.

Recommendations to Decrease/Limit Patient Radiation Exposure in Myocardial Perfusion and Cardiac FDG PET


 

Are there steps physicians can take to reduce the risk of CIN?



Yes. SCAI's Catheterization Laboratory Standards Committee has analyzed the available data about CIN and issued recommendations for the prevention of CIN at each stage in the treatment of patients who are undergoing interventional procedures that involve the use of contrast media. 

The first step is to assess a patient's kidney function. SCAI recommends that physicians assess kidney function using estimated glomerular filtration rate, or eGFR, which takes into account the patient's age, race, and other factors. The eGFR is most accurate assessment of kidney function currently available. 

Another important precaution for preventing CIN is to ensure that the patient is adequately hydrated before, during, and after interventional procedures that involve the use of contrast dyes. Patients who are at increased risk for CIN should come into the hospital far enough in advance of the procedure so that they can be given appropriate intravenous hydration. 

Patients should consult with their physicians to determine whether to discontinue taking non-steroidal anti-inflammatory agents (NSAIDS) 24-48 hours before a procedure. These medications include ibuprofen and Naprosyn, and are commonly taken for headaches, backaches, and arthritis pain. NSAIDS may cause a decrease in blood flow to the kidneys and, when combined with contrast media, are more likely to interfere with kidney function in some patients. 

SCAI recommends that physicians use low osmolar and iso-osmolar contrast agents instead of high-osmolar agents in patients at increased risk for CIN. Osmolarity refers to the number of ion particles in the blood. A low-osmolar agent is thinner than a higher osmolar agent and is considered to be safer because it is less likely to interfere with blood flow to the kidneys. Iso-osmolar contrast agents are even thinner. 

Patients at increased risk for CIN should be monitored closely after their procedure for any signs of kidney failure. In patients at increased risk for CIN, a serum creatinine test should be done during follow-up to provide assess kidney function. This test measures the amount of creatinine, an amino acid excreted by the kidneys, in a person's blood work. 

Future Radiation Protection During Cardiac Catheterization



Cardiac catheterization, whether for diagnostic procedures or cardiac interventions, will continue to be performed using ionizing radiation as a result of excellent temporal and spatial resolution afforded by fluoroscopy. It is probable that future improvements in detector sensitivity coupled with improved digitalization of the images will mean that radiation doses are reduced. It is apparent from the observations made concerning radiation doses following the introduction of flat-bed scanners that optimization of pulse frequency, filter settings, AEC management and collimation are also important when new technology is used.


Optimization of the system settings should be combined with the continued education of operators and ancillary staff.[12,65] The increasing complexity of digital imaging systems may distract staff from the importance of operator-dependent factors in controlling the levels of radiation during cardiac catheterization. In spite of the significant improvements in image quality and potential for radiation dose-lowering from digitalization and flat-bed panel systems, operator factors remain perhaps the most important determinants of radiation exposure during cardiac catheterization.

Cardiac CT Dose



Cardiac CT is emerging as an important tool for the diagnosis and monitoring of heart disease. The incidence of heart disease in the United States is already quite high and is expected to increase as the “baby boomer” segment of the population ages. To use complex multiple-row detector CT scanners most efficiently for cardiac examinations, it is important to understand many of the technical components. In addition, radiation dose for cardiac CT studies is fairly high and demands constant vigilance.

Cardiac CT techniques are generally similar to those used for routine chest CT—with a few important exceptions. Although the tube potential (measured in kilovolts) and current (measured in milliamperes) are often similar to those used for a routine chest CT study, the rotation speed is generally the fastest that can be achieved by the scanner model, and the pitch value used is often much lower than that used in routine chest examinations (typically 0.2–0.4). The use of extremely low pitch values can cause some alarm with respect to radiation dose values. However, the scan extent (start-to-stop distance for the image acquisition) is generally limited (usually 12 cm), which results in a dose range that is near accepted values for similar CT examinations of the torso.

Reduction of Cardiac CT Dose


The dose of radiation delivered during a cardiac CT examination can be decreased by implementing features such as tube current modulation, cardiac phase–specific scanning, and increased pitch.
For instance, all scanners are designed so that the x-ray tube current can be modulated, or varied, while the gantry spins around the patient. The tube current thus can be decreased when the beam crosses areas of the body with less tissue or lower tissue density, thereby lowering the total overall dose of radiation needed (9). In general CT examinations, dose reductions on the order of 20%–40% have been achieved with these approaches (10–12).

A dose-reduction approach specific to cardiac CT is to decrease the x-ray tube current during phases of the cardiac cycle that are expected to contain fairly large amounts of motion. Images acquired during these phases are expected to be of lower quality and of less value for interpretation; thus, the radiologist may choose that they not be captured for use in image reconstruction. Dose reductions of up to 50% have been achieved with this feature (13). This option may or may not be successful at individual facilities; many radiologists appear to demand images from all cardiac cycles to review regardless of their temporal resolution quality.

A third means of reducing the effective dose is to use a higher pitch value. This method is particularly used for selective patients with higher-than-normal heart rates, as with the new dual x-ray tube CT scanner discussed earlier (14). The overall dose of radiation delivered can also be substantially decreased with this method.

Combining two or more of these dose reduction options during a cardiac CT study may result in even greater decreases in the dose of radiation delivered.

Children are more vulnerable for CELL PHONE RADIATION



According to two studies completed in France and Germany, small children absorb twice as much radiation when using cell phones as adults do, due to their small ears, and thinner skin and skulls.

The French report, completed for TELECOM ParisTech by Dr. Joe Wiart, found young children can have radiation exposures twice that of adults, possibly even higher.

The second study, conducted by Niels Kuster, director of the IT’IS Foundation, had the same results. Completed for the German Federal Office of Radiation Protection (BfS), the study concluded that children’s eyes and bone marrow received more exposure.

The studies' findings support a 1996 study by Prof. Om Gandhi at the University of Utah, which found that radiation penetrates the brain of a child 4.2 times faster than an adult brain, and a child's eye more than 12 times faster.

What is radiation?



Radiation is energy that travels in the form of waves or high speed particles.

When we hear the word ' radiation,' we generally think of nuclear power plants, nuclear weapons, or radiation treatments for cancer. We would also be correct to add 'microwaves, radar, electrical power lines, cellular phones, and sunshine' to the list. There are many different types of radiation that have a range of energy forming an electromagnetic spectrum. However, when you see the word 'radiation' on this web site, we are referring to the types of radiation used in nuclear power, nuclear weapons, and medicine. These types of radiation have enough energy to break chemical bonds in molecules or remove tightly bound electrons from atoms, thus creating charged molecules or atoms (ions). These types of radiation are referred to as 'ionizing radiation.'

What is radioactivity?

Radioactivity is the property of some atoms that causes them to spontaneously give off energy as particles or rays. Radioactive atoms emit ionizing radiation when they decay.

CT Dose Modulation


Many of the dose-reduction strategies described previously result in a trade-off of image quality. For example, if tube current or voltage is reduced, the radiation dose is reduced, but so is image quality. CT manufacturers have tried to come up with new technologies that can reduce dose without significantly compromising imaging quality. One of these techniques is dose modulation.

Dose modulation is a technique by which the CT scanner modifies the tube current in response to the patient's attenuation, to maintain the same image quality for the least possible tube current. Dose modulation can be performed in the z-axis where tube current changes along the length of the patient, in the xy-plane (angular modification), or can be a combination of the 2 (xyz–dose modulation). Recent studies demonstrated that dose modulation is capable of providing a reduction in radiation dose without significant image compromise, including for neuroradiology CT protocols, where up to 60% dose reduction was achieved for noncontrast CT of the brain in adult and pediatric patients; for CT studies of the cervical spine; and for CTA studies.

Implementation of dose modulation requires a team effort between radiologists, technicians, and physicists, who have to select a preset noise index (NI) that describes the level of noise acceptable to the radiologist for a given CT examination. The CT scanner then automatically selects, within a preset range, the tube current (milliampere) required to maintain the level of noise under the noise index, taking into consideration the patient's attenuation. Identification of optimal signal intensity-to-noise ratio for each type of CT protocol requires fine-tuning to lower the milliamperes as much as possible while preserving image quality. When we transitioned to dose modulation, we initially set the NI at a low value (as recommended by the manufacturer) and progressively increased it until the image quality was deemed insufficient. This decision was a consensus decision by the 9 faculty members of our neuroradiology section.